Determination of Fano factor and pre-amplifier noise from the measurement of energy resolution of a HPGe detector
A HPGe detector and γ-ray multinuclide standard solution were used to get several data pairs of photon energy E and detector’s energy resolution FWHM. A series of graph of FWHM2 versus E was plotted to determine the detector’s Fano factor and pre-amplifier noise, which yielded the mean values of 0.1...
Main Authors: | Samat, S.B., Priharti, W. |
---|---|
Format: | Article |
Language: | English |
Published: |
Universiti Kebangsaan Malaysia
2015
|
Online Access: | http://journalarticle.ukm.my/8953/ http://journalarticle.ukm.my/8953/ http://journalarticle.ukm.my/8953/1/17_W._Priharti_.pdf |