Determination of Fano factor and pre-amplifier noise from the measurement of energy resolution of a HPGe detector

A HPGe detector and γ-ray multinuclide standard solution were used to get several data pairs of photon energy E and detector’s energy resolution FWHM. A series of graph of FWHM2 versus E was plotted to determine the detector’s Fano factor and pre-amplifier noise, which yielded the mean values of 0.1...

Full description

Bibliographic Details
Main Authors: Samat, S.B., Priharti, W.
Format: Article
Language:English
Published: Universiti Kebangsaan Malaysia 2015
Online Access:http://journalarticle.ukm.my/8953/
http://journalarticle.ukm.my/8953/
http://journalarticle.ukm.my/8953/1/17_W._Priharti_.pdf