Determination of peak- to-total ratio versus gamma-ray energy of a HPGe detector

This note describes a method for the determination of the peak-to-total ratio, P/T versus gamma-ray energy of range 59.54-1836.03 keV of a HPGe detector. Results show that this ratio decreases with energy. To check the validity of these results, two approaches of results comparison, namely with the...

Full description

Bibliographic Details
Main Authors: Supian Samat, Lee , Yuen Kong
Format: Article
Published: Universiti Kebangsaan Malaysia 2006
Online Access:http://journalarticle.ukm.my/3975/
http://journalarticle.ukm.my/3975/
id ukm-3975
recordtype eprints
spelling ukm-39752012-03-28T01:02:46Z http://journalarticle.ukm.my/3975/ Determination of peak- to-total ratio versus gamma-ray energy of a HPGe detector Supian Samat , Lee , Yuen Kong This note describes a method for the determination of the peak-to-total ratio, P/T versus gamma-ray energy of range 59.54-1836.03 keV of a HPGe detector. Results show that this ratio decreases with energy. To check the validity of these results, two approaches of results comparison, namely with the ratio reported by Cesana and Terrani in year 1989 and with the shape of total efficiency reported by Debertin and Helmer in year 1988 were taken. lt is shown that this method gives satisfactory results since the obtained values of P/T and the shape of the total efficiency curve were in good agreement with the reported works. The precautional step that need to be taken when dealing with energy 59.54 keV 241Am source and the reason for taking the total efficiency as the evaluation tool are given. Universiti Kebangsaan Malaysia 2006-07 Article PeerReviewed Supian Samat , and Lee , Yuen Kong (2006) Determination of peak- to-total ratio versus gamma-ray energy of a HPGe detector. Sains Malaysiana, 35 (1). pp. 57-60. ISSN 0126-6039 http://www.ukm.my/jsm/english_journals/vol35num1_2006/vol35num1_06page57-60.html
repository_type Digital Repository
institution_category Local University
institution Universiti Kebangasaan Malaysia
building UKM Institutional Repository
collection Online Access
description This note describes a method for the determination of the peak-to-total ratio, P/T versus gamma-ray energy of range 59.54-1836.03 keV of a HPGe detector. Results show that this ratio decreases with energy. To check the validity of these results, two approaches of results comparison, namely with the ratio reported by Cesana and Terrani in year 1989 and with the shape of total efficiency reported by Debertin and Helmer in year 1988 were taken. lt is shown that this method gives satisfactory results since the obtained values of P/T and the shape of the total efficiency curve were in good agreement with the reported works. The precautional step that need to be taken when dealing with energy 59.54 keV 241Am source and the reason for taking the total efficiency as the evaluation tool are given.
format Article
author Supian Samat ,
Lee , Yuen Kong
spellingShingle Supian Samat ,
Lee , Yuen Kong
Determination of peak- to-total ratio versus gamma-ray energy of a HPGe detector
author_facet Supian Samat ,
Lee , Yuen Kong
author_sort Supian Samat ,
title Determination of peak- to-total ratio versus gamma-ray energy of a HPGe detector
title_short Determination of peak- to-total ratio versus gamma-ray energy of a HPGe detector
title_full Determination of peak- to-total ratio versus gamma-ray energy of a HPGe detector
title_fullStr Determination of peak- to-total ratio versus gamma-ray energy of a HPGe detector
title_full_unstemmed Determination of peak- to-total ratio versus gamma-ray energy of a HPGe detector
title_sort determination of peak- to-total ratio versus gamma-ray energy of a hpge detector
publisher Universiti Kebangsaan Malaysia
publishDate 2006
url http://journalarticle.ukm.my/3975/
http://journalarticle.ukm.my/3975/
first_indexed 2023-09-18T19:40:18Z
last_indexed 2023-09-18T19:40:18Z
_version_ 1777405545001517056