Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions

The aims of this study were to compare the entrance surface dose (ESD) between anteroposterior (AP) supine with posteroanterior (PA) prone projection of computed radiography (CR) abdominal examination and to determine the relationship between body mass index (BMI) and ESD of a patient. AP supine and...

Full description

Bibliographic Details
Main Authors: Abdul Aziz Ismail, Mazlyfarina Mohamad, Rozilawati Ahmad, Ahmad Bazlie Abdul Kadir
Format: Article
Language:English
Published: Penerbit Universiti Kebangsaan Malaysia 2017
Online Access:http://journalarticle.ukm.my/11489/
http://journalarticle.ukm.my/11489/
http://journalarticle.ukm.my/11489/1/17216-56278-1-PB.pdf
id ukm-11489
recordtype eprints
spelling ukm-114892018-04-02T02:17:41Z http://journalarticle.ukm.my/11489/ Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions Abdul Aziz Ismail, Mazlyfarina Mohamad, Rozilawati Ahmad, Ahmad Bazlie Abdul Kadir, The aims of this study were to compare the entrance surface dose (ESD) between anteroposterior (AP) supine with posteroanterior (PA) prone projection of computed radiography (CR) abdominal examination and to determine the relationship between body mass index (BMI) and ESD of a patient. AP supine and PA prone projections on the same patients for CR abdominal examination of intravenous urography (IVU) were acquired on 50 patients at Hospital Raja Permaisuri Bainun, Ipoh. All the radiographic examinations were carried out on a Siemens Multixtop general x-ray unit and the images were processed with CR Carestream Direct view Max. Entrance surface dose (ESD) in miligray (mGy) was measured using optical stimulated luminescence dosimeters (OSLD) calibrated by the Malaysian Nuclear Agency. Data were analyzed using dependent t-test comparing the AP and PA projections on the same subject and Pearson correlation was used to determine the relationship between BMI and percentage of reduction of ESD. Results showed a significant different (p < 0.01) between AP supine (mean ESD = 6.42 ± 7.13 mGy) and PA prone (mean ESD = 3.92 ± 3.56 mGy) projection at all BMI. The BMI has a positive correlation with percentage of reduction of ESD (r = 0. 61) and was statistically significant (p < 0.01). In conclusion, PA abdomen prone projection significantly reduces the radiation dose and there is a positive correlation between BMI and percentage of reduction of ESD. The use of PA prone projection for CR abdominal examination should be considered as the routine projection at all BMIs level. Awareness that as the BMI increases the ESD also increases at a moderate positive linear relationship. Penerbit Universiti Kebangsaan Malaysia 2017 Article PeerReviewed application/pdf en http://journalarticle.ukm.my/11489/1/17216-56278-1-PB.pdf Abdul Aziz Ismail, and Mazlyfarina Mohamad, and Rozilawati Ahmad, and Ahmad Bazlie Abdul Kadir, (2017) Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions. Jurnal Sains Kesihatan Malaysia, 15 (2(SI)). pp. 173-178. ISSN 1675-8161 http://ejournal.ukm.my/jskm/issue/view/894
repository_type Digital Repository
institution_category Local University
institution Universiti Kebangasaan Malaysia
building UKM Institutional Repository
collection Online Access
language English
description The aims of this study were to compare the entrance surface dose (ESD) between anteroposterior (AP) supine with posteroanterior (PA) prone projection of computed radiography (CR) abdominal examination and to determine the relationship between body mass index (BMI) and ESD of a patient. AP supine and PA prone projections on the same patients for CR abdominal examination of intravenous urography (IVU) were acquired on 50 patients at Hospital Raja Permaisuri Bainun, Ipoh. All the radiographic examinations were carried out on a Siemens Multixtop general x-ray unit and the images were processed with CR Carestream Direct view Max. Entrance surface dose (ESD) in miligray (mGy) was measured using optical stimulated luminescence dosimeters (OSLD) calibrated by the Malaysian Nuclear Agency. Data were analyzed using dependent t-test comparing the AP and PA projections on the same subject and Pearson correlation was used to determine the relationship between BMI and percentage of reduction of ESD. Results showed a significant different (p < 0.01) between AP supine (mean ESD = 6.42 ± 7.13 mGy) and PA prone (mean ESD = 3.92 ± 3.56 mGy) projection at all BMI. The BMI has a positive correlation with percentage of reduction of ESD (r = 0. 61) and was statistically significant (p < 0.01). In conclusion, PA abdomen prone projection significantly reduces the radiation dose and there is a positive correlation between BMI and percentage of reduction of ESD. The use of PA prone projection for CR abdominal examination should be considered as the routine projection at all BMIs level. Awareness that as the BMI increases the ESD also increases at a moderate positive linear relationship.
format Article
author Abdul Aziz Ismail,
Mazlyfarina Mohamad,
Rozilawati Ahmad,
Ahmad Bazlie Abdul Kadir,
spellingShingle Abdul Aziz Ismail,
Mazlyfarina Mohamad,
Rozilawati Ahmad,
Ahmad Bazlie Abdul Kadir,
Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions
author_facet Abdul Aziz Ismail,
Mazlyfarina Mohamad,
Rozilawati Ahmad,
Ahmad Bazlie Abdul Kadir,
author_sort Abdul Aziz Ismail,
title Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions
title_short Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions
title_full Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions
title_fullStr Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions
title_full_unstemmed Radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions
title_sort radiation dose obtained from abdominal computed radiography: comparison between supine and prone positions
publisher Penerbit Universiti Kebangsaan Malaysia
publishDate 2017
url http://journalarticle.ukm.my/11489/
http://journalarticle.ukm.my/11489/
http://journalarticle.ukm.my/11489/1/17216-56278-1-PB.pdf
first_indexed 2023-09-18T20:00:28Z
last_indexed 2023-09-18T20:00:28Z
_version_ 1777406814624677888