Development of portable digital spirometer using NI sbRIO
One of the effective ways to diagnose various respiratory diseases is using spirometry test. Good spirometer comes with excellent graphical user interface. Spirometer is used to measure lung parameters such as FVC (Forced Vital Capacity) and FEV1 (Forced Expiratory Volume in the first second). T...
Main Authors: | , , , |
---|---|
Format: | Conference or Workshop Item |
Language: | English English |
Published: |
Institute of Electrical and Electronics Engineers Inc.
2018
|
Subjects: | |
Online Access: | http://irep.iium.edu.my/67315/ http://irep.iium.edu.my/67315/ http://irep.iium.edu.my/67315/ http://irep.iium.edu.my/67315/1/67315_Development%20of%20portable%20digital%20spirometer%20using%20NI%20sbRIO_complete.pdf http://irep.iium.edu.my/67315/2/67315_Development%20of%20portable%20digital%20spirometer%20using%20NI%20sbRIO_scopus.pdf |
id |
iium-67315 |
---|---|
recordtype |
eprints |
spelling |
iium-673152018-11-01T08:04:47Z http://irep.iium.edu.my/67315/ Development of portable digital spirometer using NI sbRIO Ibrahim, Siti Noorjannah Jusoh, Ahmad Zamani Malik, N. Abd Mazalan, S. TK7800 Electronics. Computer engineering. Computer hardware. Photoelectronic devices TK7885 Computer engineering One of the effective ways to diagnose various respiratory diseases is using spirometry test. Good spirometer comes with excellent graphical user interface. Spirometer is used to measure lung parameters such as FVC (Forced Vital Capacity) and FEV1 (Forced Expiratory Volume in the first second). The aim of this project is to develop a portable spirometer using NI sbRIO FPGA board and LabVIEW Software. The LabVIEW program consists of two parts: 1) the data acquisition and 2) the disease diagnosis. The data acquisition part will obtain data from user in terms of breathed air. Meanwhile, the disease diagnosis part will use the obtained data to analyze the disease. The system uses National Instrument Single Board Rio 9636 (NI sbRIO) for digital data conversion and data acquisition and breathing hardware from which users will used to flow air into the spirometer. The breathing hardware uses Venturi’s principle where the difference of the two pressures with different tube diameters used to measure flow rate and volume of air exhaled over a period of time; to determine the respiratory conditions. Results obtained from three different volunteers with different health performances are also presented in this paper. The FEV1 ratio of a healthy volunteer is 81.1%, an asthma volunteer is 72.04%, while a volunteer with suspected bronchitis is 33.4%. Based on these results, unhealthy users tend to have smaller value of FEV1 with lower area under the curve when compared to a healthy user. The test results are represented in parameters of pulmonary functions which include Volume-time and Flowvolume graphs. Institute of Electrical and Electronics Engineers Inc. 2018-03-09 Conference or Workshop Item PeerReviewed application/pdf en http://irep.iium.edu.my/67315/1/67315_Development%20of%20portable%20digital%20spirometer%20using%20NI%20sbRIO_complete.pdf application/pdf en http://irep.iium.edu.my/67315/2/67315_Development%20of%20portable%20digital%20spirometer%20using%20NI%20sbRIO_scopus.pdf Ibrahim, Siti Noorjannah and Jusoh, Ahmad Zamani and Malik, N. Abd and Mazalan, S. (2018) Development of portable digital spirometer using NI sbRIO. In: 4th IEEE International Conference on Smart Instrumentation, Measurement and Applications, ICSIMA 2017, 28 - 30 November 2017, Putrajaya. https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=8311987 10.1109/ICSIMA.2017.8311987 |
repository_type |
Digital Repository |
institution_category |
Local University |
institution |
International Islamic University Malaysia |
building |
IIUM Repository |
collection |
Online Access |
language |
English English |
topic |
TK7800 Electronics. Computer engineering. Computer hardware. Photoelectronic devices TK7885 Computer engineering |
spellingShingle |
TK7800 Electronics. Computer engineering. Computer hardware. Photoelectronic devices TK7885 Computer engineering Ibrahim, Siti Noorjannah Jusoh, Ahmad Zamani Malik, N. Abd Mazalan, S. Development of portable digital spirometer using NI sbRIO |
description |
One of the effective ways to diagnose various
respiratory diseases is using spirometry test. Good spirometer
comes with excellent graphical user interface. Spirometer is used
to measure lung parameters such as FVC (Forced Vital Capacity)
and FEV1 (Forced Expiratory Volume in the first second). The
aim of this project is to develop a portable spirometer using NI
sbRIO FPGA board and LabVIEW Software. The LabVIEW
program consists of two parts: 1) the data acquisition and 2) the
disease diagnosis. The data acquisition part will obtain data from
user in terms of breathed air. Meanwhile, the disease diagnosis
part will use the obtained data to analyze the disease. The system
uses National Instrument Single Board Rio 9636 (NI sbRIO) for
digital data conversion and data acquisition and breathing
hardware from which users will used to flow air into the
spirometer. The breathing hardware uses Venturi’s principle
where the difference of the two pressures with different tube
diameters used to measure flow rate and volume of air exhaled
over a period of time; to determine the respiratory conditions.
Results obtained from three different volunteers with different
health performances are also presented in this paper. The FEV1
ratio of a healthy volunteer is 81.1%, an asthma volunteer is
72.04%, while a volunteer with suspected bronchitis is 33.4%.
Based on these results, unhealthy users tend to have smaller
value of FEV1 with lower area under the curve when compared
to a healthy user. The test results are represented in parameters
of pulmonary functions which include Volume-time and Flowvolume
graphs. |
format |
Conference or Workshop Item |
author |
Ibrahim, Siti Noorjannah Jusoh, Ahmad Zamani Malik, N. Abd Mazalan, S. |
author_facet |
Ibrahim, Siti Noorjannah Jusoh, Ahmad Zamani Malik, N. Abd Mazalan, S. |
author_sort |
Ibrahim, Siti Noorjannah |
title |
Development of portable digital spirometer using NI sbRIO |
title_short |
Development of portable digital spirometer using NI sbRIO |
title_full |
Development of portable digital spirometer using NI sbRIO |
title_fullStr |
Development of portable digital spirometer using NI sbRIO |
title_full_unstemmed |
Development of portable digital spirometer using NI sbRIO |
title_sort |
development of portable digital spirometer using ni sbrio |
publisher |
Institute of Electrical and Electronics Engineers Inc. |
publishDate |
2018 |
url |
http://irep.iium.edu.my/67315/ http://irep.iium.edu.my/67315/ http://irep.iium.edu.my/67315/ http://irep.iium.edu.my/67315/1/67315_Development%20of%20portable%20digital%20spirometer%20using%20NI%20sbRIO_complete.pdf http://irep.iium.edu.my/67315/2/67315_Development%20of%20portable%20digital%20spirometer%20using%20NI%20sbRIO_scopus.pdf |
first_indexed |
2023-09-18T21:35:35Z |
last_indexed |
2023-09-18T21:35:35Z |
_version_ |
1777412798864687104 |