A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography

Introduction: Phantom studies facilitate the implementation of radiation dose surveillance as a function of radiographic technical parameters for minimizing patient radiation dose. The evidence of such investigations can then be used to evaluate technical parameters used in the radiographic procedur...

Full description

Bibliographic Details
Main Authors: Moey, Soo Foon, Ahmad Shazli, Zubir
Format: Article
Language:English
English
Published: Iranian Journal of Medical Physics 2018
Subjects:
Online Access:http://irep.iium.edu.my/66511/
http://irep.iium.edu.my/66511/
http://irep.iium.edu.my/66511/
http://irep.iium.edu.my/66511/1/66511_A%20phantom%20study%20for%20the%20optimization%20of%20image%20quality.pdf
http://irep.iium.edu.my/66511/2/66511_A%20phantom%20study%20for%20the%20optimization_SCOPUS.pdf
id iium-66511
recordtype eprints
spelling iium-665112018-12-31T09:34:37Z http://irep.iium.edu.my/66511/ A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography Moey, Soo Foon Ahmad Shazli, Zubir QC Physics R Medicine (General) Introduction: Phantom studies facilitate the implementation of radiation dose surveillance as a function of radiographic technical parameters for minimizing patient radiation dose. The evidence of such investigations can then be used to evaluate technical parameters used in the radiographic procedures to reduce radiation dose without compromising the image quality. Material and Methods: This experimental study was carried out using an anthropomorphic phantom and the Leeds test object. Computed radiographic system was utilized and the images were printed for objective evaluation. Dose-area-product (DAP) readings were obtained using a DAP meter for the technical parameters employed for the radiographic procedures. Results: The use of 0.2 mm additional copper filtration resulted in the lowest radiation doses for all four radiographic procedures (i.e. posteroanterior chest, anteroposterior abdomen and lumbar sacral spine projections). The highest tube potential appropriate to the body part being imaged, patient size, image receptor response and required information resulted in the minimum radiation dose to the patient without compromising the image quality. The focus to film distance utilized for the radiographic procedure must be in accordance with the focus to grid distance specified by the manufacturer when using the bucky to eliminate grid “cut-off.” Conclusion: The optimization of image quality and radiation dose can be accomplished by using a phantom and selecting the imaging parameters that yield an acceptable image quality with the lowest entrance surface dose while considering the adjustment for patient size. Iranian Journal of Medical Physics 2018-10-01 Article PeerReviewed application/pdf en http://irep.iium.edu.my/66511/1/66511_A%20phantom%20study%20for%20the%20optimization%20of%20image%20quality.pdf application/pdf en http://irep.iium.edu.my/66511/2/66511_A%20phantom%20study%20for%20the%20optimization_SCOPUS.pdf Moey, Soo Foon and Ahmad Shazli, Zubir (2018) A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography. Iranian Journal of Medical Physics, 15 (4). pp. 271-276. ISSN 1735-160X http://ijmp.mums.ac.ir/article_10573_727fdf069f10be3ab4b4327471d44b3f.pdf 10.22038/IJMP.2018.29825.1325
repository_type Digital Repository
institution_category Local University
institution International Islamic University Malaysia
building IIUM Repository
collection Online Access
language English
English
topic QC Physics
R Medicine (General)
spellingShingle QC Physics
R Medicine (General)
Moey, Soo Foon
Ahmad Shazli, Zubir
A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography
description Introduction: Phantom studies facilitate the implementation of radiation dose surveillance as a function of radiographic technical parameters for minimizing patient radiation dose. The evidence of such investigations can then be used to evaluate technical parameters used in the radiographic procedures to reduce radiation dose without compromising the image quality. Material and Methods: This experimental study was carried out using an anthropomorphic phantom and the Leeds test object. Computed radiographic system was utilized and the images were printed for objective evaluation. Dose-area-product (DAP) readings were obtained using a DAP meter for the technical parameters employed for the radiographic procedures. Results: The use of 0.2 mm additional copper filtration resulted in the lowest radiation doses for all four radiographic procedures (i.e. posteroanterior chest, anteroposterior abdomen and lumbar sacral spine projections). The highest tube potential appropriate to the body part being imaged, patient size, image receptor response and required information resulted in the minimum radiation dose to the patient without compromising the image quality. The focus to film distance utilized for the radiographic procedure must be in accordance with the focus to grid distance specified by the manufacturer when using the bucky to eliminate grid “cut-off.” Conclusion: The optimization of image quality and radiation dose can be accomplished by using a phantom and selecting the imaging parameters that yield an acceptable image quality with the lowest entrance surface dose while considering the adjustment for patient size.
format Article
author Moey, Soo Foon
Ahmad Shazli, Zubir
author_facet Moey, Soo Foon
Ahmad Shazli, Zubir
author_sort Moey, Soo Foon
title A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography
title_short A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography
title_full A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography
title_fullStr A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography
title_full_unstemmed A phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography
title_sort phantom study for the optimization of image quality and radiation dose for common radiographic examinations in digital radiography
publisher Iranian Journal of Medical Physics
publishDate 2018
url http://irep.iium.edu.my/66511/
http://irep.iium.edu.my/66511/
http://irep.iium.edu.my/66511/
http://irep.iium.edu.my/66511/1/66511_A%20phantom%20study%20for%20the%20optimization%20of%20image%20quality.pdf
http://irep.iium.edu.my/66511/2/66511_A%20phantom%20study%20for%20the%20optimization_SCOPUS.pdf
first_indexed 2023-09-18T21:34:26Z
last_indexed 2023-09-18T21:34:26Z
_version_ 1777412726105047040